INTRA ORAL SENSOR MAINTENANCE

Maintenance Instructions

  1. Remove and discard all protective hygienic barriers and / or sheaths from the Sensor prior to removing disposable gloves.
  2. Place the Sensor on a tray covered by a disposable liner, or in a receptacle that can be thoroughly disinfected.
  3. Remove and discard gloves.
  4. Wash hands and put on a new pair of disposable gloves.
  5. Disconnect the Sensor from the Remote Module.
  6. If the Sensor or cable are visibly soiled (e.g., with blood or saliva), each should be cleaned with a soapy cloth or paper towel, and then dried with a clean lint-free cloth or paper towel.
  7. Thoroughly wipe the Sensor and Sensor cable (if applicable) with the disinfecting product recommended above. Do not expose the contacts of the sensor/remote module connection to liquid.
  8. Repeat step 7. When the Sensor has been wiped two times, continue with the following steps. It is very important to follow these next two steps!
  9. Remove potential chemical build-up from the Sensor by wiping it with a sterile sponge saturated with de-ionized water.
  10. Use a sterile dry sponge to dry the Sensor and Sensor cable, as needed.
  11. Place the Sensor in a clean environment, ready for next use.
  12. Reconnect the Sensor waiting to take an image
  13. Remove and discard gloves.